Difference between revisions of "Adam Equipment CPWplus200 Configuration"
Line 25: | Line 25: | ||
|- | |- | ||
|Transmission Mode | |Transmission Mode | ||
− | |trn | + | |trn 1 |
− | | | + | |Default. Can set to trn 2 during trouble shooting. |
|- | |- | ||
|} | |} |
Latest revision as of 05:00, 15 February 2015
See Section Section 13 of the CPW200plus User Manual for detailed instructions.
With the scale off, hold the Tare/Zero button and press the On/Off button.
Parameter | Setting | Rational |
---|---|---|
Auto Power Off | Pr oFF | Don't want the scale to turn off after a period of inactivity. |
Backlight always off | bL 1 | Increases battery life when AC power fails. |
Baudrate | 9600 | Default |
8 Bit no parity | PAr 1 | Default |
Transmission Mode | trn 1 | Default. Can set to trn 2 during trouble shooting. |